Connection modes


Types of connections

The user can establish two different types of connections to a chromatography system:

  • Control mode connection: The user is able to actively control the system.

  • View mode connection: The user can monitor the system activity but cannot control the system.

Several simultaneous connections can be established to one system, but only one may be in control mode. The other connections are in view mode.


Status bar information and possible actions

The System Control module displays information in the status bar regarding the connection mode. The table below explains the status bar text and the possible actions the user can take:

Status bar text

Connection mode

Possible actions

(nothing)

Not connected

To establish a connection, either

  • click the System Connection icon

    or

  • select System:Connect.

Controlled by: <user>

Control mode

To leave the system but retain the connection with the System Control module, either

  • click Disconnect from the System icon

    or

  • select System:Disconnect.

Note: You may leave the system locked or unlocked.

Controlled by: <other user>

View mode

The indicated user has a control mode connection.

None. Even if you click the System Connection icon it has no effect.

Locked by: <other user>

View mode

The indicated user has left the system in a locked state.

Click the System Connection icon to establish a control mode connection. You must supply a password, either

  • the locking password

    or

  • your logon password (if you have Unlock locked systems access).

Note: The password is case sensitive.

System is available

View mode

A user has left the system in an unlocked state.

Click the System Connection icon to establish a control mode connection.


2005-06-15