-
Quantitation requires
that the components of interest are completely resolved from all
other components in the chromatogram. Overlapping peaks will produce unreliable
results.
-
The peak integration parameters (baseline settings)
must be correctly selected. The default settings will be satisfactory
in many cases, but the integration results have to be checked for
all chromatograms.
-
All integrations must be performed using the same
X-axis base unit. For highest reliability, time is the recommended
unit.
-
The concentration levels of the standard have to
be accurately prepared. Errors in the amount or concentration values
will lead to unpredictable results.
-
Self-imposed limitations, such as the use of a small
number of concentration levels of the standard, also limits precision.
-
Precision is improved by the appropriate choice
of the concentration range of the standard. The range should extend
across the presumed amount in the sample.
-
Use of the most appropriate curve model will maximize
precision.
-
Accuracy is improved if several runs are performed
at each level.
-
All the runs should be performed consecutively to
reduce systematic errors and thereby maximize precision.