How to change the Correction factors
|
If the
readings described in the instruction above are acceptable at both
0% and 100%, the Correction factors do
not need to be changed. If the Correction
factors do not produce an acceptable result, they must
be adjusted in the Method Editor module.
The table below describes how to change the Correction
factors:
Step
|
Action
|
1
|
Choose Edit:BufferPrep
Recipes.
Result: The BufferPrep Recipes dialog box
opens.
|
2
|
Select the recipe from the Recipe droplist
and click the Edit button.
Result: The Edit Recipe dialog box opens.
|
3
|
Click the Correction factors button.
Result: The Correction Factors dialog box
opens.
|
4
|
Example: If the
pH is set to 7.0 and the actual pH is 7.1, the Correction factor is 0.1. If the
actual pH is 6.9, the Correction factor is
-0.1.
|
Note: If there already
are Correction factors,
the measured pH deviation should be added to the old factors.
|