The standard
series should include standard concentrations that extend beyond
the lower and upper limits of the sample amount. If an internal
standard is used, the internal standard must be added in the same
concentration in all standards.
Methods created from a wizard
If the method is created from a wizard for ÄKTAdesign systems,
you may select the correct standard concentration level in the variable Quantitation_Type. You can also
set the level after the run has been performed. Each level is an
alias for a specific concentration of the standard.
The list below describes how the levels are applied:
-
Level 1 should be
selected for the standard with the highest or lowest concentration.
-
The levels must be set in consecutive order of changing
concentration of the standard.
-
All runs with the same concentration must be given
the same level.
|